SEC-logo
Certification

ISO/IEC 17025 Accredited by A2LA

SOLAR ELECTRONICS
10866 Chandler Boulevard
North Hollywood, CA 91601

CALIBRATION

Valid To: March 31, 2027
Certificate Number: 7219.01

In recognition of the successful completion of the A2LA evaluation process (including an assessment of the organization’s compliance with R205 – A2LA’s Calibration Program Requirements), accreditation is granted to this laboratory to perform the following calibrations and dimensional inspections1:

I. Electrical – RF/Microwave

Parameter/Equipment Range CMC2, 3(±) Comments
Transformers
     
Insertion Loss – AC Port:
(30 Hz to 250 kHz)
(- 7.0 to 0.1) dB 0.45 dB MIL-STD-461/462
network analyzer, (KEYSIGHT E5061B)
Insertion Loss – Secondary:
(30 Hz to 250 kHz)
(- 7.5 to - 0.4) dB 0.45 dB
Current and Injection Probes      
Insertion Loss (dB):
(20 Hz to 10 kHz)
(10 kHz to 300 MHz)
(300 kHz to 1 GHz)
(- 79.8 to - 71.5) dB
(- 71.1 to - 17.8) dB
(- 38.3 to - 13.8) dB
- 17 % - 11 dB
- 0.16 % + 0.36 dB
1.8 dB
MIL-STD-461, MIL-STD-462 Notice 3, DO-160, network analyzer, (KEYSIGHT E5061B)
Correction Factor (dB):
(20 Hz to 10 kHz)
(10 kHz to 300 MHz)
(300 kHz to 1 GHz)
(37.5 to 45.7) dB
(- 16.2 to 37.1) dB
(- 20.1 to 4.3) dB
16 % - 5.0 dB
0.16 % + 0.41 dB
1.8 dB
LISNs      
Impedance Magnitude:
1 kHz to 150 MHz
(150 to 400) MHz
400 MHz to 1 GHz
(0 to 46.9) Ω
(47.3 to 77.1) Ω
(55.6 to 78.8) Ω
0.15 % + 0.10 Ω
3.9 % - 1.7 Ω
- 10 % + 11 Ω
MIL-STD 461, MILSTD 461 (alt. 5μH LISN from Appendix A), MIL-STD-462 Notice 3, DO-160, CISPR 16-1, CISPR 16-1-2, CISPR 25, FCC Part 15, FCC Part 18, DEF STAN 59-411 Part 3, ANSI C63.4(1992), ANSI C63.4(2003), ISO 7637-2, network analyzer, (KEYSIGHT E5061B plus calibration kit 85032E)
Impedance Phase:
(100 kHz to 10 MHz)
(10 to 100) MHz
(100 MHz to 1 GHz)
(- 8.6 to 85.5) °
(- 1.0 to 8.0) °
(- 60.1 to 0.8) °
1.9 °
0.39 °
4.8 °
Insertion Loss:
(1 kHz to 1 GHz)
(- 62.2 to - 0.5) dB 0.42 dB
Correction Factor:
(1 kHz to 1 GHz)
(- 29.7 to 0.1) dB 0.68 dB
Isolation (dB):
(1 kHz to 100 MHz)
(- 70.5 to - 5.6) dB 0.42 dB
Calibration Fixtures      
VSWR – w/o Probe:
(20 Hz to 500 MHz)
(500 MHz to 1 GHz)
(1.0 to 2.1) lin
(2.2 to 5.0) lin
4.0 % - 0.03 lin
0.51 lin
MIL-STD-461, MILSTD-462 Notice 3, DO-160,network analyzer, (KEYSIGHT E5061B)
VSWR – w/ Probe:
(20 Hz to 500 MHz)
(500 MHz to 1 GHz)
(1.0 to 1.6) lin
(1.5 to 4.3) lin
1.9 % - 0.01 lin
21 % - 0.30 lin
Attenuators      
Insertion Loss:
(100 Hz to 1 GHz)
(- 41.4 to - 39.9) dB 0.37 dB MIL-STD-461, (KEYSIGHT E5061B
network analyzer, and 85032E)
VSWR Input:
(100 Hz to 1 GHz)
(1.1 to 2.9) lin 26 % - 0.26 lin
VSWR Output:
(100 Hz to 1 GHz)
(1.1 to 1.9) lin 100 % - 1.1 lin
Feedthrough Capacitors      
Insertion Loss:
(1 to 300) kHz
(300 kHz to 1 MHz)
(- 52.2 to - 5.3) dB
(- 80.0 to – 41.4) dB
0.51 dB
1.8 dB
MIL-STD 461, MILSTD-462 Notice 3, DO-160, network analyzer, (E5061B plus 85032E)
Loop Antennas and Sensors      
Correction Factor:
(10 Hz to 2 kHz)
(2 kHz to 1 MHz)
(63.5 to 80.0) dB
(38.8 to 62.9) dB
12 % - 6.8 dB
0.49 % + 0.17 dB
MIL-STD-461 (RS101, RE101), network analyzer, (KEYSIGHT E5061B)
RF Couplers      
Insertion Loss:
(10 kHz to 400 MHz)
(- 48.6 to - 38.5) dB 0.41 dB MIL-STD-461, network analyzer, (KEYSIGHT E5061B plus 85032E)
Resistive Devices      
VSWR:
(10 Hz to 1 GHz)
(1.0 to 1.4) lin 18 % - 0.17 lin MIL-STD 461, network analyzer, (KEYSIGHT E5061B plus 85032E)

1 This laboratory offers commercial calibration service.

2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

3 In the statement of CMC, percentages are to be read as percentage of reading, unless otherwise noted.